Welcome

Welcome to the website of Clean Room Validation Ltd. Clean Room Validation Ltd, based in Glasgow, provides an independent specialist clean room testing and validation service.

Our Aim

To ensure your clean air environment operates to the specified ISO classification and design specification, and to help you, our customer, maintain regulatory compliance and protect your product and your operators from contamination.

Table 1

Schedule of testing to demonstrate compliance with particle concentration limits (ISO 14644-2).

Classification Maximum time interval Test method

≤ ISO Class 5

6 months

Annex B in ISO 14644-1:1999

≥ ISO Class 5

12 months

Annex B in ISO 14644-1:1999

Notes

Particle counts will normally be performed in the operational state, but may also be performed in the at-rest state in accordance with the designated ISO classification.

Table 2

Schedule of additional tests for all classes (ISO 14644-2).

Test Parameter Maximum time interval Test procedure

Airflow volume1 or airflow velocity

12 months

ISO 14644-3:- clause B.4

Air pressure difference2

12 months

ISO 14644-3:- clause B.5

Notes

These tests may normally be performed in either the operational state or the at-rest state in accordance with the designated ISO classification.

1 Airflow volume may be determined by either velocity or volume measurement techniques.

2 This test will not apply to clean zones which are not totally enclosed.

Table 3

Schedule of optional tests (ISO 14644-2).

Test Parameter Class Suggested maxmimum time interval Test procedure

Installed filter leakage

All classes

24 months

ISO 14644-3:- clause B.6

Airflow visualisation

All classes

24 months

ISO 14644-3:- clause B.7

Recovery

All classes

24 months

ISO 14644-3:- clause B.13

Containment leakage

All classes

24 months

ISO 14644-3:- clause B.14